Message on display Start mass out of range / Install weighing pan RADWAG MA 5Y

Start mass out of range

At power-up the analyzer compares the weight of the pan and handle with the stored start mass. If it is out of range, the message asks you to install the pan.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn the device off.
  2. Install the pan correctly and leave it empty.
  3. Turn the device on again.

Possible causes

  • The weighing pan is not installed
  • Sample on the pan at power-up

Do you need service?

If the message remains with an empty, correctly fitted pan, the start mass must be checked by a technician.

Source: RADWAG official user manual · Open manufacturer document