Calibration failed
The calibration process could not be completed. The balance continues to operate with the old calibration values.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Use a calibration weight of the correct value.
- Reduce air drafts and vibration.
- Check that the security switch is not locked and repeat the calibration.
- Do not touch the sealed security switch on legal-for-trade approved devices.
Possible causes
- Wrong calibration weight
- Unstable environment
- The security switch is in the locked position
Do you need service?
If calibration still cannot be done with the correct weight under suitable conditions, contact technical service.
Models showing this code: SPX
Source: OHAUS official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.