Code on display Err 8.2 OHAUS Scout SPX

Below power-on zero range

The weight measured at switch-on is below the power-on zero range. This is generally seen because the weighing pan is not installed.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check that the weighing pan is installed and seated correctly.
  2. Switch the balance off and on again.

Possible causes

  • The weighing pan is not installed
  • The weighing pan is seated incorrectly
  • Calibration drift

Do you need service?

If the error persists even with the weighing pan installed correctly, contact technical service.

Models showing this code: SPX

Source: OHAUS official user manual