Above power-on zero range
The weight on the weighing pan at switch-on exceeds the power-on zero range. The balance cannot take its initial zero with this load.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty the weighing pan.
- Check that nothing is touching the weighing pan.
- Switch the balance off and on again.
Possible causes
- There was a load on the weighing pan at switch-on
- An object is touching the weighing pan
- Calibration drift
Do you need service?
If the error persists at switch-on with an empty weighing pan, contact technical service to check the calibration or load cell.
Models showing this code: SPX
Source: OHAUS official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.