Code on display Err 8.3 OHAUS Navigator NVT

Overload

The reading is above capacity plus 9 divisions. The balance cannot weigh this load.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Remove the load from the pan.
  2. Split the load into portions within capacity.
  3. Check that nothing else touches the pan.
  • A load above capacity can permanently damage the load cell.

Possible causes

  • A load heavier than the capacity
  • An object pressing on the pan

Do you need service?

If Err 8.3 appears with an empty pan, the load cell must be checked by a technician.

Models showing this code: NVT6201, NVT10201

Source: OHAUS official user manual · Open manufacturer document