Piece weight too small
The average piece weight is below the smallest step the balance can show, so counting is not possible. The balance shows the error and leaves the mode or goes to the clear APW screen (Clr.APW).
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Take the reference again with more pieces.
- Clear the piece weight (Clr.APW) and restart counting.
- Use a balance with finer readability for very light parts.
Possible causes
- Pieces being too light
- Reference was taken incorrectly
Do you need service?
This message relates to the piece weight and does not indicate a fault in the balance.
Models showing this code: NVT6201, NVT10201
Source: OHAUS official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.