Code on display Lo ref OHAUS Navigator NVT

Reference weight too low

The balance cannot calculate a reliable average piece weight from so little weight. Counting does not start until a valid reference is set.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Place more reference pieces on the pan.
  2. Check that you placed the number of pieces selected for the reference.
  3. Repeat the reference step.

Possible causes

  • Too few pieces placed for the reference
  • Pieces being too light

Do you need service?

No technician is needed. The message clears once the reference weight is increased.

Models showing this code: NVT6201, NVT10201

Source: OHAUS official user manual · Open manufacturer document