Code on display Err 11.0 OHAUS MB23 / MB25 / MB27

AC power unstable

The heater cannot be set to the correct temperature without a stable supply. The manual recommends using a power line conditioner.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Plug the unit into another socket.
  2. Switch off high-power devices on the same line.
  3. Supply the analyzer through a power line conditioner.

Possible causes

  • Mains voltage fluctuations
  • High-power devices running on the same line

Do you need service?

If Err 11.0 appears with a stable supply, the power board must be checked by a technician.

Models showing this code: MB23, MB25, MB27

Source: OHAUS official user manual · Open manufacturer document