Code on display Err 8.4 OHAUS MB23 / MB25 / MB27

Pan support removed during weighing

The analyzer lost the weight of the pan support during weighing. The measurement cannot continue.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Re-install the pan support.
  2. Check that the pan support is seated correctly.
  3. Clean the area around the pan support.
  • The heater and sample chamber may be very hot. Take care when opening the cover.

Possible causes

  • Pan support removed during weighing
  • Pan support moved out of place
  • Dirt or residue inside the unit

Do you need service?

If Err 8.4 appears with the pan support in place, the unit must be cleaned and checked by a technician.

Models showing this code: MB23, MB25, MB27

Source: OHAUS official user manual · Open manufacturer document