Code on display Err 8.2 OHAUS MB23 / MB25 / MB27

Pan support not ınstalled

The analyzer does not see the expected weight of the pan support at start-up. The service manual notes that a dirty pan support can also cause this error.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Turn the device off.
  2. Install the pan support correctly.
  3. Check that the pan support is clean.
  4. Turn the device on again.
  • The heater and sample chamber may be very hot. Take care when opening the cover.

Possible causes

  • Pan support not installed
  • Pan support seated wrongly or dirty
  • Load cell fault

Do you need service?

If Err 8.2 appears with the pan support correctly installed, the load cell must be checked by a technician.

Models showing this code: MB23, MB25, MB27