Code on display Err 10.7 OHAUS MB23 / MB25 / MB27

Temperature out of control

According to the service manual, the temperature has risen 20 °C above the target. Unstable mains voltage, a sensor fault or a board fault can cause this.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Stop the test and let the analyzer cool down.
  2. Separate the analyzer from high-power devices on the same outlet.
  • The heater and sample chamber may be very hot. Take care when opening the cover.
  • Do not leave the analyzer unattended while the temperature is out of control.

Possible causes

  • Unstable mains voltage
  • Temperature sensor failure
  • Main board fault

Do you need service?

If the error remains, the temperature sensor and main board must be checked by a technician.

Models showing this code: MB23, MB25, MB27