Code on display Err 10.3 OHAUS MB23 / MB25 / MB27

Temperature sensor out of range, high

The analyzer cannot read the temperature sensor within its valid range. Heating cannot be controlled safely in this state.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Switch the analyzer off and let it cool down.
  2. Switch it on again and check whether the error remains.
  • The heater and sample chamber may be very hot. Take care when opening the cover.
  • Do not open the housing. Internal parts carry mains voltage.

Possible causes

  • Sensor cable connection problem
  • Temperature sensor failure
  • Main board fault

Do you need service?

The sensor cables, temperature sensor and main board must be checked by a technician.

Models showing this code: MB23, MB25, MB27

Source: OHAUS official user manual · Open manufacturer document