Code on display REF.WT Err OHAUS Defender 3000 / i-DT33

Reference weight too small

The reference is not saved because the average piece weight is below 0.1 division. Heavier parts or a more sensitive scale are needed.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Increase the number of pieces in the reference sample.
  2. Use parts whose average piece weight is at least 0.1 division.

Possible causes

  • Piece weight too small relative to the scale resolution
  • Insufficient reference sample

Do you need service?

This is an application limit and does not require service.