Code on display Inf 03 Minebea Intec Signum 1/2/3

Calibration not completed in time

If the device has not warmed up enough or the reading does not stabilize, the adjustment times out. The manual recommends waiting for the device to warm up again and repeating the adjustment.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Leave the device on for a while and wait for it to warm up.
  2. Eliminate sources of vibration and drafts.
  3. Repeat the calibration.

Possible causes

  • Insufficient warm-up time
  • Unstable environment (vibration, drafts)

Do you need service?

If it recurs under suitable conditions, a service check is required.

Models showing this code: Signum 1, Signum 2, Signum 3, SIWABBP, SIWSBBP, SIWSBBS, SIWRDCP, SIWADCP, SIWSDCP, SIWSDCS

Source: Minebea Intec official user manual