Code on display Err 341 Minebea Intec Signum 1/2/3

RAM data lost

According to the manual, the cause is a discharged internal battery that preserves the data. The device must stay connected to power for at least 10 hours to charge the battery.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Leave the device connected to power for at least 10 hours.
  2. Check the date/time and settings, and re-enter them if necessary.

Possible causes

  • Internal backup battery discharged
  • Device left without power for a long time

Do you need service?

If the error recurs despite the long charge, contact technical service to replace the internal battery.

Models showing this code: Signum 1, Signum 2, Signum 3, SIWABBP, SIWSBBP, SIWSBBS, SIWRDCP, SIWADCP, SIWSDCP, SIWSDCS

Source: Minebea Intec official user manual