Mettler Toledo

IND246 error codes

24 error codes

Review the 24 error and warning messages from the Mettler Toledo IND246 manufacturer manual, with meanings and fix steps. Device type: industrial weighing terminal.

Covered models: IND246, IND246 POWERCELL

Err 0001 Metrology switch off An approval region is selected but the W&M (metrology) switch SW1-1 is off. Err 0002 Setting cannot be changed while approved This setup parameter cannot be changed while the terminal is in approved status. Err 0003 Division value will be reset Reports that the increment (division) value will be reset because the capacity was changed. Err 0004 Range 2 capacity invalid In multi-range mode, the range 2 capacity must be greater than range 1. Err 0005 Capacity limit exceeded The entered capacity value exceeds the permitted limit. Err 0006 Calibration weight out of range The test weight entered for calibration is outside the permitted range. Err 0007 GEO value out of limits The entered GEO (geographic adjustment) value is outside the permitted range. Err 0008 Invalid capacity value The entered capacity value is not accepted by the terminal. Err 0009 Range division ratio incorrect In multi-range mode, the ratio of the range 2 division to the range 1 division must be less than 50. E E E Zero could not be captured at startup With startup zeroing enabled, the weight on the platform is outside the startup zeroing range. exceeds zero range (+) Above zero range The ZERO key was pressed but the weight is above the zeroing range. illegal scale mode Zeroing not possible in net mode The ZERO key was pressed while the terminal was in net (tared) mode. motion Command not completed due to motion The zero, tare or print command could not be completed because motion was detected on the scale. no demand output Print connection not defined Printing was requested but no port has a demand output defined. scale not ready Not ready to print Printing was requested but the print lock or the auto print reset condition was not met. scale over capacity Print attempted over capacity Printing was requested while the scale was over capacity. scale under zero Print attempted below zero Printing was requested while the weight was below zero (display blank). tare not enabled Tare disabled The TARE key was pressed but the tare function is turned off in the settings. tare too small Tare value too small The TARE key was pressed when the weight was less than one division. tare under zero Tare below zero The TARE key was pressed while the weight was below zero. tare exceeds limit Preset tare greater than capacity The entered preset tare value is greater than the scale capacity. keypad tare disabled Keypad tare off Preset tare entry was attempted but keypad tare is turned off in the settings. PDX CELL xx NO RESPONSE POWERCELL cell not responding In the POWERCELL version, load cell number xx is not responding to the terminal. PDX CELL xx NOT FOUND POWERCELL cell not found Load cell number xx could not be detected on the POWERCELL network.