110% of full scale exceeded
When a load far above system capacity is detected, the transmitter raises an alarm, the relays open and the analog output drops to its lowest level.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Unload the excess load.
- Check that the weighed structure does not touch its surroundings (pipes, platform).
- Verify that the alarm clears once the load returns to normal.
- Do not load the load cells above their rated capacity.
Possible causes
- Silo/tank overfilled
- Mechanical binding or side force on the load cell
- Incorrect full scale setting
Do you need service?
If the alarm persists after the load is removed, the full scale setting or the load cell must be checked by a technician.
Models showing this code: TLB, TLB485, TLB CANOPEN, TLB DEVICENET, TLB PROFI, TLB PROFINET IO, TLB ETHERNET/IP, TLB MODBUS/TCP, TLB ETHERCAT
Source: Laumas official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.