Load cell connection error
According to the manual, this alarm also appears when the A/D converter fails, when the EX-/REF- and EX+/REF+ jumpers are not fitted on a 4-wire cell, or when the reference leads are connected incorrectly. When the alarm occurs, the relays open and the analog output drops to its lowest value.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Check the load cell cable for visible breaks, crushing, or moisture.
- Visually check that the terminals in the junction box are not loose.
- Disconnect the power and reconnect it.
- Do not disconnect terminal connections while power is on, and leave wiring work to an authorized person.
Possible causes
- Broken or loose load cell cable
- Missing jumper on a 4-wire cell (EX–REF)
- Excessive load cell signal
- Converter fault
Do you need service?
If the alarm persists although the cables look intact, the cell connection, jumpers, or converter must be measured by a qualified technician.
Models showing this code: TLB, TLB485, TLB CANOPEN, TLB DEVICENET, TLB PROFI, TLB PROFINET IO, TLB ETHERNET/IP, TLB MODBUS/TCP, TLB ETHERCAT
Source: Laumas official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.