Code on display ErCEL Laumas TLB

Load cell connection error

According to the manual, this alarm also appears when the A/D converter fails, when the EX-/REF- and EX+/REF+ jumpers are not fitted on a 4-wire cell, or when the reference leads are connected incorrectly. When the alarm occurs, the relays open and the analog output drops to its lowest value.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check the load cell cable for visible breaks, crushing, or moisture.
  2. Visually check that the terminals in the junction box are not loose.
  3. Disconnect the power and reconnect it.
  • Do not disconnect terminal connections while power is on, and leave wiring work to an authorized person.

Possible causes

  • Broken or loose load cell cable
  • Missing jumper on a 4-wire cell (EX–REF)
  • Excessive load cell signal
  • Converter fault

Do you need service?

If the alarm persists although the cables look intact, the cell connection, jumpers, or converter must be measured by a qualified technician.

Models showing this code: TLB, TLB485, TLB CANOPEN, TLB DEVICENET, TLB PROFI, TLB PROFINET IO, TLB ETHERNET/IP, TLB MODBUS/TCP, TLB ETHERCAT

Source: Laumas official user manual