Unload pan for automatic adjustment
ACJ models adjust themselves with the internal weight when the temperature changes or about every four hours. This cannot be done with a loaded pan, so the balance waits for the pan to be emptied.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Remove the load from the pan.
- Close the draft shield doors and avoid vibration.
- Wait until adjustment ends and the gram display returns.
- To avoid interrupting a measurement, press ON/OFF when PLS.CAL appears to postpone adjustment by two minutes.
Possible causes
- Sample or container on the pan when adjustment was due
- Room temperature changed by about 2 °C
- About four hours since the last adjustment
Do you need service?
Automatic adjustment runs by itself once the pan is empty.
Models showing this code: ACJ 80-4M, ACJ 100-4M, ACJ 200-4M, ACJ 300-4M
Source: KERN & Sohn official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.