Code on display PLS.CAL KERN & Sohn ACJ

Unload pan for automatic adjustment

ACJ models adjust themselves with the internal weight when the temperature changes or about every four hours. This cannot be done with a loaded pan, so the balance waits for the pan to be emptied.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Remove the load from the pan.
  2. Close the draft shield doors and avoid vibration.
  3. Wait until adjustment ends and the gram display returns.
  4. To avoid interrupting a measurement, press ON/OFF when PLS.CAL appears to postpone adjustment by two minutes.

Possible causes

  • Sample or container on the pan when adjustment was due
  • Room temperature changed by about 2 °C
  • About four hours since the last adjustment

Do you need service?

Automatic adjustment runs by itself once the pan is empty.

Models showing this code: ACJ 80-4M, ACJ 100-4M, ACJ 200-4M, ACJ 300-4M

Source: KERN & Sohn official user manual · Open manufacturer document