Code on display ERR C KERN & Sohn ACJ

Zero shift during adjustment

At the start of adjustment the balance reads a zero value with an empty pan. If this value is far from the expected one, adjustment is cancelled.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Press ON/OFF to return to weighing mode.
  2. Empty the pan and check that it is installed correctly.
  3. Restart the adjustment with the CAL key.

Possible causes

  • Objects on the pan when adjustment starts
  • The weighing pan is not installed
  • The weighing pan touching something

Do you need service?

If ERR C repeats with an empty, correctly fitted pan, the weighing cell must be checked by a technician.

Models showing this code: ACJ 80-4M, ACJ 100-4M, ACJ 200-4M, ACJ 300-4M

Source: KERN & Sohn official user manual · Open manufacturer document