Code on display BUSY KERN & Sohn ACJ

Pan loaded at adjustment start

Adjustment must start with an empty pan. When a load is detected, the balance waits and asks for it to be removed.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Remove the object from the pan.
  2. Continue the adjustment.

Possible causes

  • Sample or container left on the pan
  • Dust or small particles on the pan

Do you need service?

Adjustment continues once the pan is empty.

Models showing this code: ACJ 80-4M, ACJ 100-4M, ACJ 200-4M, ACJ 300-4M