Initial zero point exceeded
At startup the value read from the load cell is too far from the zero point set at calibration. The scale cannot obtain a reliable zero in this state, so it does not start weighing. This is usually caused by an object left on or under the weighing pan and rarely by a load cell fault.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Turn off the scale and remove everything from the weighing pan.
- Check under and around the weighing pan and clear any trapped objects.
- Make sure the weighing pan is seated correctly, then turn the scale on again with an empty pan.
- Do not strike the load cell or force the weighing pan.
Possible causes
- A load was on the weighing pan at switch-on
- Foreign object trapped under the weighing pan
- Load cell fault or damage
Do you need service?
If the error persists with an empty pan, the manual recommends service to check/replace the load cell.
Models showing this code: JCL
Source: Jadever official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.