A/D resolution range exceeded
The scale cannot calculate weight because the analog signal is above or below the limits the converter can read. In this case the manual points to an A/D board or load cell fault.
This fault needs a technician.
What should you do?
You can safely do these while waiting for service.
- Empty the weighing pan, turn the scale off and turn it on again after a few seconds.
- Do not open the device housing. Board and load cell replacement is a job for technical service.
Possible causes
- A/D converter (measurement board) fault
- Load cell fault
- Load cell connection problem
Do you need service?
The A/D board or load cell must be tested and replaced, so technical service is required.
Models showing this code: JCL
Source: Jadever official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.