Code on display Err50 Esit Smart-2

Calibration switch is off

While P2 is shorted, settings can be viewed but not saved, and the calibration menu cannot be entered. This is a lock that protects the settings for legal metrology.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Give up changing the setting and return to the main screen.
  2. If a setting change is needed, contact technical service.
  • On sealed devices, do not open the seal or the P2 switch yourself.

Possible causes

  • P2 calibration switch is in the locked position
  • Attempt to change settings on a sealed device

Do you need service?

Access to the calibration switch requires breaking the seal, so the work must be done by technical service.

Models showing this code: Smart-2

Source: Esit official user manual