Code on display Coeff.OutofLimit Esit ECI

Calibration coefficient out of limit

A valid calibration coefficient could not be obtained because the test load does not match the value entered. The manual recommends increasing the load or lowering the value to be calibrated.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Increase the calibration load.
  2. Check that the calibration value entered is the same as the actual load on the platform.
  3. Repeat the calibration.
  • Leave calibration of sealed scales to technical service.

Possible causes

  • Calibration load is insufficient
  • Calibration value entered is too high
  • Load cell connection problem

Do you need service?

If the error occurs even with the correct load, service calibration is required.

Models showing this code: ECI

Source: Esit official user manual