Printer function mismatch
The printer type that is set does not support subtotal/grand total output. According to the manual, DLP label printers cannot print totals.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Check which printer model you are using.
- If you use a DEP printer, set F40 to "2".
- If you use a DLP label printer, print single items instead of totals.
Possible causes
- Requesting a total printout with a DLP label printer
- Wrong F40 printer type setting
Do you need service?
If the error persists with the correct printer type set, the printer connection must be checked by service or the dealer.
Models showing this code: CPS-PLUS
Source: CAS official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.