Aczet scale error codes

2 series · 0 error codes

Aczet, formerly Citizen Scale (I), is an India-based manufacturer of laboratory and jewelry scales, producing analytical, precision and carat balances and moisture analyzers. This guide covers the error messages of the CY series analytical/precision balances and CY-C series laboratory balances, as published in Aczet's own manuals.

--OL-- Overload CY Series · --OL-- indicates that the scale has detected a load exceeding its capacity. --UL-- Underload / pan missing CY Series · --UL-- indicates that the scale has detected a load below zero. Error 1 Reference weight too low CY Series · Error 1 indicates that the reference weight placed for piece counting, percent, special unit, or check weighing is too low. Error 2 Pan loaded at startup CY Series · Error 2 indicates that the load on the pan during calibration is more than 10% of capacity. Error 3 Calibration weight error CY Series · Error 3 indicates that the weight was not placed in time during calibration or is out of tolerance. Error 4 GLP footer not printed CY Series · Error 4 indicates an attempt to enter the user menu without printing the footer while GLP is on. Error 6 Calibration when not at zero CY Series · Error 6 indicates an attempt to start calibration while the display shows a value other than 0.00. Error 7 Invalid date / time CY Series · Error 7 indicates that the entered date or time value is invalid. Error 8 Print mode not suitable for GLP CY Series · Error 8 indicates an attempt to turn on GLP while the print option is AUTO, AUTO LOAD, or CONTINUOUS. Error 9 Real-time clock error CY Series · Error 9 indicates that the balance's internal real-time clock (RTC) is not working. Error29 Calibration and clock error CY Series · Error 29 indicates that a calibration error and a real-time clock (RTC) error occurred together. Error39 Calibration weight and clock error CY Series · Error 39 indicates that a calibration weight error and a real-time clock (RTC) error occurred together. -Err3- Outside tare range CY-C Series · Err3 indicates that the value to be tared is outside the allowed tare range. -Err8- Tare / zeroing timeout CY-C Series · Err8 indicates that the taring or zeroing operation could not be completed within the allowed time. -NULL- Zero value from converter CY-C Series · NULL indicates that the balance is receiving a zero value from its measuring transducer. -FULL- Measuring range exceeded CY-C Series · FULL indicates that the load on the weighing pan exceeds the balance's measuring range. -LH- Initial mass error CY-C Series · LH indicates that the initial mass measured by the balance at startup is not within the expected range.