Initial mass error
At startup the balance compares the weight of the pan and its holder with the stored initial mass. If the difference exceeds the permitted range, LH is displayed. Typical causes are a load on the pan or a pan that is not installed. According to the manual, on two-pan models LH also appears if the balance is turned on with both pans installed.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Turn the balance off and empty the pan.
- Check that the weighing pan is installed correctly.
- On a two-pan model, make sure only one pan is installed.
- Turn the balance back on.
Possible causes
- Load on the weighing pan at startup
- The weighing pan is not installed or is seated incorrectly
- Both pans installed at the same time on a two-pan model
Do you need service?
If LH persists with an empty, correctly installed pan, the initial mass setting or the measuring system must be checked by technical service.
Models showing this code: CY-165C, CY-205C, CY-265C
Source: Aczet official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.