Code on display Error 9 Aczet CY Series

Real-time clock error

The clock circuit that keeps date and time is not working, so a correct timestamp cannot be generated on GLP printouts. In this case the manual says to contact technical service.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Turn the balance off and on again. If the error persists, contact service.
  • Do not open the device to replace the backup battery.

Possible causes

  • RTC circuit fault
  • Clock backup battery depleted

Do you need service?

Because the clock circuit is on the board, its repair must be done by technical service.

Models showing this code: CY 124, CY 224, CY 123, CY 223, CY 1202, CY 2202, CY 3102, CY 1003, CY 253K, CY 1503K

Source: Aczet official user manual