Underload / pan missing
According to the manual, the most common cause is that the weighing pan is not mounted on the weighing spindle or is seated incorrectly. In this case the scale cannot see its expected tare weight.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Mount the weighing pan on the weighing spindle.
- Check that the pan is seated correctly and flat.
- Turn the scale off and on again with the pan empty.
Possible causes
- The weighing pan is not installed
- Pan seated incorrectly on the spindle
Do you need service?
If --UL-- persists with the pan installed correctly, contact technical service to have the measuring system checked.
Models showing this code: CY 124, CY 224, CY 123, CY 223, CY 1202, CY 2202, CY 3102, CY 1003, CY 253K, CY 1503K
Source: Aczet official user manual
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.