Code on display --UL-- Aczet CY Series

Underload / pan missing

According to the manual, the most common cause is that the weighing pan is not mounted on the weighing spindle or is seated incorrectly. In this case the scale cannot see its expected tare weight.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Mount the weighing pan on the weighing spindle.
  2. Check that the pan is seated correctly and flat.
  3. Turn the scale off and on again with the pan empty.

Possible causes

  • The weighing pan is not installed
  • Pan seated incorrectly on the spindle

Do you need service?

If --UL-- persists with the pan installed correctly, contact technical service to have the measuring system checked.

Models showing this code: CY 124, CY 224, CY 123, CY 223, CY 1202, CY 2202, CY 3102, CY 1003, CY 253K, CY 1503K

Source: Aczet official user manual