Code on display -Err3- Aczet CY-C Series

Outside tare range

According to the manual, negative values cannot be tared. Pressing the tare key with a negative display gives Err3. In this case the manual recommends zeroing first and then repeating the tare operation.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Empty the pan and zero the display.
  2. Place the container on the pan and tare once the display is stable.

Possible causes

  • Tare taken with a negative display
  • Tare attempted after the load was removed from the pan

Do you need service?

The error clears when the tare operation is repeated after zeroing. No service is needed.

Models showing this code: CY-165C, CY-205C, CY-265C

Source: Aczet official user manual