Calibration weight too heavy
In calibration with an external weight, the weight placed on the weighing pan measured too heavy compared with the expected calibration value. The balance does not accept the calibration to prevent an incorrect adjustment.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Press the CAL key to return to weighing mode.
- Check that the weighing pan and pan support are installed correctly.
- Verify that the weight you are using matches the calibration value on the display.
- Repeat the calibration with the correct weight.
Possible causes
- Using a weight heavier than the calibration value on the display
- Extra load or foreign object on the weighing pan
- Weighing pan or pan support installed incorrectly
Do you need service?
If CAL E continues with the correct weight and a correctly installed weighing pan, technical service must check the balance's adjustment and load cell.
Models showing this code: HR-AZ
Source: A&D official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.