Code on display Error7 A&D HR-AZ

Internal weight mechanism error

The mechanism that engages the internal calibration weight did not work properly. In this case calibration with the internal weight (one-touch) cannot be completed and the balance's accuracy cannot be updated.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Empty the weighing pan and check that it is seated correctly.
  2. Turn the balance off and on and start the weighing process from the beginning.
  3. Repeat the calibration in an environment free of vibration without touching the weighing pan.
  • Do not open the device to reach the internal weight mechanism.

Possible causes

  • Vibration or shock during calibration
  • Touching the weighing pan during calibration or leaving a load on it
  • Mechanical fault in the internal weight mechanism

Do you need service?

If the error repeats, contact technical service for a mechanical check of the internal calibration weight mechanism.

Models showing this code: HR-AZ