Internal weight mechanism error
The mechanism that engages the internal calibration weight did not work properly. In this case calibration with the internal weight (one-touch) cannot be completed and the balance's accuracy cannot be updated.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty the weighing pan and check that it is seated correctly.
- Turn the balance off and on and start the weighing process from the beginning.
- Repeat the calibration in an environment free of vibration without touching the weighing pan.
- Do not open the device to reach the internal weight mechanism.
Possible causes
- Vibration or shock during calibration
- Touching the weighing pan during calibration or leaving a load on it
- Mechanical fault in the internal weight mechanism
Do you need service?
If the error repeats, contact technical service for a mechanical check of the internal calibration weight mechanism.
Models showing this code: HR-AZ
Source: A&D official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.