Code on display -E A&D HR-AZ

Weighing pan / underload error

The balance detects a lower weight than should be on the weighing pan. This usually indicates that the weighing pan or pan support is not installed or is not fully seated.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check the weighing pan and pan support and place them correctly.
  2. Press the ON:OFF key twice to return to weighing mode.
  3. If the error continues, calibrate the balance.

Possible causes

  • Weighing pan not installed
  • Weighing pan or pan support seated incorrectly
  • Balance calibration drift

Do you need service?

If -E continues even with the weighing pan installed correctly and calibration done, contact technical service for a load cell check.

Models showing this code: HR-AZ

Source: A&D official user manual · Open manufacturer document