Code on display CAL E A&D GH

Calibration weight too heavy

The balance cannot match the weight on the pan to the selected calibration value. The calibration is not saved.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check the value of the calibration weight.
  2. Make sure only the weight is on the pan.
  3. Press the CAL key to return to weighing mode and calibrate again.

Possible causes

  • Wrong nominal weight used
  • Calibration weight value set incorrectly
  • Extra object on the pan

Do you need service?

If CAL E appears with the correct weight, the balance adjustment must be checked by a technician.

Models showing this code: GH-120, GH-200, GH-300, GH-202, GH-252

Source: A&D official user manual · Open manufacturer document