A&D

GH error codes

25 error codes

Review the 25 error and warning messages from the A&D GH manufacturer manual, with meanings and fix steps. Device type: analytical balance.

Covered models: GH-120, GH-200, GH-300, GH-202, GH-252

Error 1 Weighing not stable Error 1 (EC,E11) means the balance cannot stabilize because of the environment. Error 6 Internal weight error Error 6 (EC,E16) means applying the internal mass did not give the expected change. Error 7 Internal mass mechanism error Error 7 (EC,E17) means the mechanism that applies the internal mass is not working properly. CAL E Calibration weight too heavy CAL E (EC,E20) means the weight placed during external calibration is heavier than expected. -CAL E Calibration weight too light -CAL E (EC,E21) means the weight placed during external calibration is lighter than expected. E Overload E means the sample on the pan is beyond the weighing capacity of the balance. -E Weighing pan error -E means the reading is too light and the weighing pan may not be installed correctly. Lo Sample too light Lo means the sample to be stored for counting or percent mode is too light. 25 PCS More pieces needed 25 PCS means the unit mass in counting mode is too light and the sample should be increased to 25 pieces. 50 PCS More pieces needed 50 PCS means the unit mass in counting mode is too light and the sample should be increased to 50 pieces. 100 PCS More pieces needed 100 PCS means the unit mass in counting mode is too light and the sample should be increased to 100 pieces. CH 0 Pan not empty, adjustment failed CH 0 means the automatic response adjustment could not run because there is something on the pan. CH nG Unstable value, adjustment failed CH nG means the automatic response adjustment could not run because the reading is unstable. rtc PF Clock battery depleted rtc PF means the backup battery for the clock has run out. Lb Low supply voltage Lb reports that the supply voltage to the balance has dropped. FUL Memory full FUL means the data memory of the balance is full. Clr Memory type mismatch Clr means the memory type set in the function table differs from the type of data stored. Err Memory data error Err means the data stored in memory has been destroyed. EC,E00 Communication protocol error EC,E00 means a protocol error occurred in communications in computer or printer communication. EC,E01 Undefined command EC,E01 means an undefined command was received in computer or printer communication. EC,E02 Balance not ready EC,E02 means a received command cannot be processed at this time in computer or printer communication. EC,E03 Command timeout EC,E03 means the next character of a command did not arrive within the time limit in computer or printer co… EC,E04 Too many characters EC,E04 means the command contained too many characters in computer or printer communication. EC,E06 Command format error EC,E06 means the command contains incorrect data in computer or printer communication. EC,E07 Parameter out of range EC,E07 means the received data exceeds the range the balance can accept in computer or printer communication.