Code on display Error 1 A&D GH

Weighing not stable

The value keeps moving, so the balance cannot complete the measurement. The manual lists vibration, drafts, temperature changes, static electricity and magnetic fields as causes.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check around the pan and make sure nothing touches it.
  2. Close the draft shield doors.
  3. Move the balance to a stable table without drafts.
  4. Press the CAL key to return to weighing mode.

Possible causes

  • Vibration or air drafts
  • Sudden temperature change
  • Static electricity or magnetic fields
  • Something touching the pan

Do you need service?

If Error 1 remains in a suitable environment, the weighing mechanism must be checked by a technician.

Models showing this code: GH-120, GH-200, GH-300, GH-202, GH-252

Source: A&D official user manual · Open manufacturer document