Weighing not stable
The value keeps moving, so the balance cannot complete the measurement. The manual lists vibration, drafts, temperature changes, static electricity and magnetic fields as causes.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Check around the pan and make sure nothing touches it.
- Close the draft shield doors.
- Move the balance to a stable table without drafts.
- Press the CAL key to return to weighing mode.
Possible causes
- Vibration or air drafts
- Sudden temperature change
- Static electricity or magnetic fields
- Something touching the pan
Do you need service?
If Error 1 remains in a suitable environment, the weighing mechanism must be checked by a technician.
Models showing this code: GH-120, GH-200, GH-300, GH-202, GH-252
Source: A&D official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.