Code on display EC,E04 A&D GH

Too many characters

The balance received a longer command than expected and does not process it. Extra characters may have been added at the end.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Check the command sent.
  2. Remove the extra characters and send the command again.

Possible causes

  • Extra characters added to the command
  • Wrong line ending setting

Do you need service?

EC,E04 is a communication setting error. It clears once the settings are corrected and no technician is needed.

Models showing this code: GH-120, GH-200, GH-300, GH-202, GH-252