Code on display EC,E00 A&D GH

Communication protocol error

The settings between the computer or printer and the balance may not match. The manual asks you to check the format, baud rate and parity.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Compare the baud rate on the balance and the connected device.
  2. Match the parity and data format settings.
  3. Send the command again.

Possible causes

  • Different baud rate or parity
  • Wrong data format
  • Cable or connection problem

Do you need service?

EC,E00 is a communication setting error. It clears once the settings are corrected and no technician is needed.

Models showing this code: GH-120, GH-200, GH-300, GH-202, GH-252

Source: A&D official user manual · Open manufacturer document