Code on display SD Error A&D GF-L

ECL repeatability error

In the self check performed with the electronically controlled load (ECL), the standard deviation of repeatability exceeded 50 divisions (50 d). This indicates that the installation environment is disturbing measurement stability.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Place the balance on a solid, vibration-free table away from drafts.
  2. Allow enough time for the balance to warm up.
  3. Run the self check again.

Possible causes

  • Vibration or air drafts
  • Balance standing on an unstable surface
  • Temperature change

Do you need service?

If SD Error persists in the self check even after the environment is corrected, the balance's weighing performance should be checked by service.

Models showing this code: GF-L

Source: A&D official user manual · Open manufacturer document