Overload
The load on the weighing pan exceeds the balance's weighing capacity. Because overload can damage the load cell (sensor), the load must be removed immediately.
You can usually fix this yourself.
What should you do?
Try the steps in order.
- Remove the sample from the weighing pan immediately.
- Check the balance's capacity (the Max value on the model label) and keep the load below the capacity.
- Check that the display returns to zero when the pan is empty.
- Do not place a load above capacity on the pan and do not drop the load onto it.
Possible causes
- A sample above capacity was placed
- Too much load placed on the pan together with the tare container
- Impact or sudden load applied to the pan
Do you need service?
If the E indication does not clear after the load is removed, the sensor may be damaged. In this case, contact technical service.
Models showing this code: GF-L
Source: A&D official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.