Code on display MW Error A&D GF-L

Minimum weight (reference value) error

An error occurred in the automatic minimum weight (reference value) determination with ECL. The minimum weight value could not be set reliably, so the operation is not completed.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Check the installation environment (vibration, drafts, temperature).
  2. Check the settings in section '6-2. Self check function / automatic setting of minimum weight with ECL' of the manual.
  3. Run the operation again.

Possible causes

  • Instability caused by the environment
  • Unsuitable minimum weight settings

Do you need service?

If MW Error persists even though the environment and settings are suitable, a service check is required.

Models showing this code: GF-L

Source: A&D official user manual · Open manufacturer document