Internal weight mechanism error
Shown on the display as Error 7: the mechanism that loads and removes the internal calibration weight is not working properly. Calibration with the internal weight cannot be performed in this case.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Empty the weighing pan and restart the procedure from the beginning.
- Turn the balance off and on again, then retry internal calibration.
- Do not open the device to interfere with the mechanism.
Possible causes
- Internal weight mechanism jamming
- Mechanism problem after transport
- Shaking during calibration
Do you need service?
If Error 7 persists when the procedure is repeated, the internal weight mechanism is faulty and must be repaired by technical service.
Models showing this code: BM
Source: A&D official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.


