Code on display Error 1 A&D BM

Stability error

This error, also shown on the display as Error 1, indicates that the weighing value could not be stabilized, so operations such as zeroing and calibration cannot be performed. The cause is usually environmental: vibration, air currents, temperature changes, static electricity, or a magnetic field.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Check around the weighing pan and make sure it does not touch the draft shield or any other object.
  2. Move the balance to a sturdy, vibration-free table away from air currents.
  3. Take the necessary precautions for statically charged or magnetic samples (e.g. an anti-static device, a suitable container).
  4. Press the CAL key to return to weighing mode and repeat the procedure.

Possible causes

  • Vibration or shaking
  • Air currents (air conditioner, door, window)
  • Static electricity or a magnetic sample
  • The weighing pan touching something

Do you need service?

According to the manual, the cause is the environmental conditions. If the error keeps recurring after the environment is corrected, a service check of the balance may be needed.

Models showing this code: BM