Code on display Error 6 A&D BM

Internal calibration weight error

Shown on the display as Error 6: when the internal calibration weight is applied, the expected change in the measured value did not occur. Internal calibration cannot be completed in this case.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Make sure the weighing pan is completely empty.
  2. Keep the balance in a vibration-free environment and restart internal calibration from the beginning.
  • Do not open the device to access the internal weight mechanism.

Possible causes

  • Load on the weighing pan during calibration
  • Vibration or shock during calibration
  • Problem in the internal weight mechanism

Do you need service?

If Error 6 appears again in repeated attempts with an empty pan, the internal weight mechanism must be checked by service.

Models showing this code: BM

Source: A&D official user manual · Open manufacturer document