Code on display OL Shimadzu UX

Overload

The balance detected more load than it can measure. The troubleshooting table of the manual says this can also relate to the pan and setup.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Reduce the load on the weighing pan.
  2. Place the pan correctly.
  3. Check that the pan supporter caps are installed.
  • Do not place loads above capacity on the pan.

Possible causes

  • Load above capacity
  • Pan displaced
  • Pan supporter caps not installed

Do you need service?

If OL continues with an empty, correctly installed pan, the weighing mechanism must be checked by a technician.

Source: Shimadzu official user manual