Code on display CHE x Shimadzu UX

Balance failure

A short countdown from CHE 5 to CHE 0 at power-on is normal. If the display stops at a CHE x message, the balance stops working because of a fault.

This fault needs a technician.

What should you do?

You can safely do these while waiting for service.

  1. Turn the balance off and unplug it.
  2. Wait one minute and turn it on again.
  3. If the message continues, note the code on the display.

Possible causes

  • Electronics board failure
  • Weighing mechanism fault

Do you need service?

CHE x is a hardware fault, the balance must be repaired by a technician.

Source: Shimadzu official user manual