Code on display OL / -OL Shimadzu TX

Weighing range error

The weighing range was exceeded. OL indicates that the capacity was exceeded, and -OL indicates that the reading has dropped far below zero. According to the manual, a displaced weighing pan can also cause this message.

You can usually fix this yourself.

What should you do?

Try the steps in order.

  1. Remove the load from the weighing pan.
  2. Place the weighing pan on the balance correctly.
  3. On models with a large weighing pan, check that the weighing pan support heads are installed.
  4. Use the balance within the specified capacity.
  • Do not place a load above capacity on the weighing pan. The load cell may be damaged.

Possible causes

  • Capacity exceeded
  • Weighing pan displaced or not installed
  • Weighing pan support heads removed (models with a large weighing pan)

Do you need service?

The message clears when the weighing pan is installed correctly and the load is reduced within capacity. No service is needed.

Models showing this code: TX

Source: Shimadzu official user manual · Open manufacturer document