Calibration error
Span calibration could not be completed. According to the manual, a large drift in the zero point or sensitivity, a container left on the pan, a displaced pan, or a wrong weight causes this error. It also occurs if the procedure is not continued within 60 seconds.
Check it yourself first. If it continues, it needs service.
What should you do?
Try the steps in order.
- Press the relevant key to return to weighing mode.
- Check that the weighing pan is installed correctly and empty.
- Turn the device off and on, then repeat span calibration by placing the correct weight in the center of the weighing pan.
- Complete the calibration steps within 60 seconds.
Possible causes
- Container or load left on the weighing pan
- Weighing pan shifted out of place
- Using the wrong calibration weight
- Large drift in the zero point or sensitivity
Do you need service?
If ERR C persists in repeated calibration with the correct weight and an empty pan, the balance's sensitivity must be checked by service.
Models showing this code: TX
Source: Shimadzu official user manual · Open manufacturer document
Legal notice
The error code explanations and checks on this page are based on manufacturer manuals and are provided for general information only; they do not replace technical service, diagnosis or warranty. The meaning of a code may vary by model, firmware version and configuration.
The decision to carry out any suggested step, and its consequences, rest with the user. Terazibul cannot be held liable for device damage, weighing errors, data loss, commercial loss or any other direct or indirect damage arising from the use of this information.
On legal-for-trade (verified) scales, adjustment and calibration must only be performed by a qualified technical service in line with legal metrology rules. Disconnect power before opening the device and contact a service provider if in doubt. Brand and model names belong to their respective owners and are used only to identify the device.