Code on display ERR C Shimadzu TW

Calibration error

Span calibration could not be completed. According to the manual, a large drift in the zero point or sensitivity, a container left on the pan, a displaced pan, or a wrong weight causes this error. It also occurs if the procedure is not continued within 60 seconds.

Check it yourself first. If it continues, it needs service.

What should you do?

Try the steps in order.

  1. Press the relevant key to return to weighing mode.
  2. Check that the weighing pan is installed correctly and empty.
  3. Turn the device off and on, then repeat span calibration by placing the correct weight in the center of the weighing pan.
  4. Complete the calibration steps within 60 seconds.

Possible causes

  • Container or load left on the weighing pan
  • Weighing pan shifted out of place
  • Using the wrong calibration weight
  • Large drift in the zero point or sensitivity

Do you need service?

If ERR C persists in repeated calibration with the correct weight and an empty pan, the balance's sensitivity must be checked by service.

Models showing this code: TW

Source: Shimadzu official user manual · Open manufacturer document